Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Enforcement
- Recall Number
- Z-1300-2026
- Event ID
- 98159
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Maquet Cardiopulmonary Gmbh
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 11, 2026
- Initiation Date
- January 9, 2026
- Classification Date
- February 5, 2026
- Address
- Kehler Str. 31, N/A, Rastatt, N/A, N/A, Germany
Description
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Model Number: 701055720; UDI-DI: 04037691816432; Serial Numbers: All units impacted;
Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of (South Korea), Latvia, Lithuania, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, and Vietnam
3050 units