FDA Enforcement
Class II
Terminated
RayStation 11B
Recall: Z-1299-2022
·
Reported July 6, 2022
Enforcement
- Recall Number
- Z-1299-2022
- Event ID
- 90505
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- RAYSEARCH LABORATORIES AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 6, 2022
- Initiation Date
- June 10, 2022
- Classification Date
- June 28, 2022
- Termination Date
- June 11, 2024
- Address
- Sveavagen 9, Stockholm, N/A, Sweden
Description
RayStation 11B
Reason
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
Code Info
UDI-DI 0735000201042620211208 Software version numbers 12.0.0.932, 12.1.0.1221, 12.0.3.68
Distribution
US Nationwide distribution in the states of NJ and TX.
Quantity
26