Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)
Enforcement
- Recall Number
- Z-1298-2022
- Event ID
- 90358
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Haemonetics Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 6, 2022
- Initiation Date
- April 19, 2022
- Classification Date
- June 28, 2022
- Termination Date
- March 27, 2024
- Address
- 125 Summer St, Boston, MA, 02110-1616, United States
Description
Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
This correction is applicable when TEG 5000 PlateletMapping (Item # 07-014, 07-015, 07-016) is performed on a TEG 5000 analyzer (Item # 07-022, 07-033) that is connected to TEG Manager Software which includes viewing of TEG 5000 results. Code Info: UDI-DI: (01)30812747018184 Software Version: TEGM 1.1 and later, TAS v4.2.3
US Nationwide.