LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
Recall: Z-1295-2018
·
Reported April 11, 2018
Enforcement
- Recall Number
- Z-1295-2018
- Event ID
- 79279
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Magellan Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 11, 2018
- Initiation Date
- January 9, 2018
- Classification Date
- April 3, 2018
- Termination Date
- December 11, 2020
- Address
- 101 Billerica Ave Bldg 4, N/A, North Billerica, MA, 01862-1271, United States
Description
LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
Reason
The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
Code Info
Part Numbers: 70-6760, 70-6529-R, 70-6760-P, and 70-6760-R. All LeadCare II and LeadCare Plus Blood Lead Analyzers with battery boards versions B and O.
Distribution
Worldwide Distribution
Quantity
6,143