TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

Recall: Z-1294-2025 · Reported March 12, 2025

Enforcement

Recall Number
Z-1294-2025
Event ID
96274
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
TAS Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 12, 2025
Initiation Date
January 27, 2025
Classification Date
March 6, 2025
Address
9100 Conroy Windermere Rd Ste 200, Windermere, FL, 34786-8431, United States

Description

TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

Reason

Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.

Code Info

Model/UDI-DI/Lot(Expiration): T-4000/00197644217267/230403(2024/04/20); T-5000/00860011435921/240801(2025/08/28), 240901(2025/09/25); T-LAP/00860011435914, 00197644927241/240702(2025/07/08), 230906(2024/11/29), 230905(2024/11/29)

Distribution

US Nationwide distribution in the states of LA, OK, TN.

Quantity

114