FDA Enforcement Class II Terminated

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.

Recall: Z-1294-2019 · Reported May 15, 2019

Enforcement

Recall Number
Z-1294-2019
Event ID
82217
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medica Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
May 15, 2019
Initiation Date
February 6, 2019
Classification Date
May 7, 2019
Termination Date
October 2, 2020
Address
5 Oak Park Dr, N/A, Bedford, MA, 01730-1430, United States

Description

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.

Reason

Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.

Code Info

Lot #s 18066, 18133, 18235, 18317

Distribution

Worldwide distribution - US Nationwide and countries of IRAQ, FRANCE, INDIA, GEORGIA, GEORGIA, FRANCE GUATEMALA, KENYA, IRAN, ESTONIA.

Quantity

87 kits