FDA Enforcement
Class II
Ongoing
Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
Recall: Z-1294-2018
·
Reported April 11, 2018
Enforcement
- Recall Number
- Z-1294-2018
- Event ID
- 79362
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Seiler Instr. & Mfg. Co. Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 11, 2018
- Initiation Date
- January 1, 2017
- Classification Date
- April 2, 2018
- Address
- 3433 Tree Court Industrial Blvd, N/A, Saint Louis, MO, 63122-6617, United States
Description
Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
Reason
The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.
Code Info
Model Number 6 ENT
Distribution
Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.
Quantity
43 units, but 1 affected