FDA Enforcement Class II Terminated

Uterine Electromyographic Monitor - Monica Novii Wireless Patch System Product Usage: The Novii Interface is an accessory to the Novii Pod which provides a means of interfacing the wireless output of the Novii Pod to the transducer inputs of a CTG Fetal monitor via three cables. The Novii Interface enables signals collected by the Novii Pod to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected.

Recall: Z-1292-2018 · Reported April 11, 2018

Enforcement

Recall Number
Z-1292-2018
Event ID
79464
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
GE Healthcare, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 11, 2018
Initiation Date
December 15, 2017
Classification Date
April 2, 2018
Termination Date
April 21, 2020
Address
3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States

Description

Uterine Electromyographic Monitor - Monica Novii Wireless Patch System Product Usage: The Novii Interface is an accessory to the Novii Pod which provides a means of interfacing the wireless output of the Novii Pod to the transducer inputs of a CTG Fetal monitor via three cables. The Novii Interface enables signals collected by the Novii Pod to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected.

Reason

if the cables of the Monica Novii System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

Code Info

Model # - 107-PT-020 ( all units distributed from January 2016- November 2017)

Distribution

Worldwide - US Nationwide Distribution.

Quantity

1742