The HomeChoice Pro APD System is a peritoneal dialysis system. The HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
Enforcement
- Recall Number
- Z-1292-2015
- Event ID
- 66737
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Baxter Healthcare Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 25, 2015
- Initiation Date
- October 16, 2013
- Classification Date
- March 18, 2015
- Termination Date
- July 21, 2017
- Address
- 25212 W. Illinois Route 120, Round Lake, IL, 60073-9799, United States
Description
The HomeChoice Pro APD System is a peritoneal dialysis system. The HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
The occluder in affected devices 1) may have sharp edges that could cut the pneumatic tubing that connects the occluder blade to the manifold causing an unrecoverable system error, and/or 2) blade may separate from the back plate and cause Slow / No Flow alarm requiring therapy termination.
Product Code: 5C8310; Serial Numbers: 80415, 680416, 680418, 680420, 680421, 680422, 680424, 680426, 680427, 680428, 680429, 680430, 680438, 680439, 680440, 680441, 680444, 680445, 680448, 680449, 680452, 680453, 680454, 680455, 680456, 680457, 680459, 680460, 680461, 680462, 680463, 680465, 680467, 680467, 680469, 680470, 680472, 680473, 680475, 680476, 680477, 680478, 680480, 680481, 680482, 680463, 680484, 680485, 680486, 680489, 680490, 680491, 680494, 680495, 680496, 680498, 680499, 680500, 680501, 680502, 680503, 680504, 680505, 680506, 680507, 680508, 680509, 680510, 680516, 680517, 680519, 680520, 680521, 680522, 680524, 680525, 680526, 680527, 680527, 680528, 680529, 680530, 680532, 680533, 680534, 680536, 680537, 680538, 680539, 680541, 680542, 680543, 680547, 680552, 680553, 680544, 680555, 680561, 680562, 680566, 680567, 680585, 680586, 680587, 680588, 680589, 680590, 680591, 680593, 680594, 680595, 680596, 680597, 680598, 680599, 680600, 680601, 680602, 680603, 680604, 680605, 680606, 680607, 680608, 680609, 680611, 680612, 680613, 680614, 680615, 680616, 680617, 680618, 680619, 680620, 680621, 680622, 680623, 680624, 680625, 680626, 680627, 680628, 680629, 680630, 680631, 680632, 680633, 680634
Worldwide Distribution-US (nationwide) including the states of AK, AR, AZ, CA, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, VT, WA, and WI, and the countries of Australia, Canada, and Mexico.
75 units (US)