FDA Enforcement Class II Ongoing

DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement

Recall: Z-1291-2025 · Reported March 12, 2025

Enforcement

Recall Number
Z-1291-2025
Event ID
96261
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
DeRoyal Industries Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 12, 2025
Initiation Date
January 31, 2025
Classification Date
March 5, 2025
Address
200 Debusk Ln, Powell, TN, 37849-4703, United States

Description

DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement

Reason

Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.

Code Info

71-C5100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578738 Lot Number: CNWKD04-04 71-C6100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578851 Lot Numbers: CNWKD04-04 CNWKD04-05 CNWKD04-10 CNWKD09-08 71-C7100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578967 Lot Numbers: CNWKD04-06 CNWKD04-07

Distribution

U.S.: IL, MN, NC, PA, and VA O.U.S.: N/A

Quantity

714 cases