BARD MARQUEE Disposable Core Biopsy Instruments XX.Xcm, XX.Xcm, XXg x XXcm REF XXXXXXX Note - for definition of the "X's refer to Tyvek Lid Label.pdf" uploaded in Associated Documents
Enforcement
- Recall Number
- Z-1289-2021
- Event ID
- 87316
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Bard Peripheral Vascular Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 31, 2021
- Initiation Date
- January 14, 2021
- Classification Date
- March 22, 2021
- Termination Date
- January 28, 2025
- Address
- 1625 W 3rd St Bldg 1, N/A, Tempe, AZ, 85281-2438, United States
Description
BARD MARQUEE Disposable Core Biopsy Instruments XX.Xcm, XX.Xcm, XXg x XXcm REF XXXXXXX Note - for definition of the "X's refer to Tyvek Lid Label.pdf" uploaded in Associated Documents
Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.
Catalog and Product Description MQ1210 BARD MARQUEE Disposable Core Biopsy Instrument, 12g x 10cm, MQ1213 BARD MARQUEE Disposable Core Biopsy Instrument, 12g x 13cm, MQ1410 BARD MARQUEE Disposable Core Biopsy Instrument, 14g x 10cm, MQ1413 BARD MARQUEE Disposable Core Biopsy Instrument, 14g x 13cm, MQ1416 BARD MARQUEE Disposable Core Biopsy Instrument, 14g x 16cm, MQ1610 BARD MARQUEE Disposable Core Biopsy Instrument, 16g x 10cm, MQ1616 BARD MARQUEE Disposable Core Biopsy Instrument, 16g x 16cm, MQ1620 BARD MARQUEE Disposable Core Biopsy Instrument, 16g x 20cm, MQ1810 BARD MARQUEE Disposable Core Biopsy Instrument, 18g x 10cm, MQ1816 BARD MARQUEE Disposable Core Biopsy Instrument, 18g x 16cm, MQ1820 BARD MARQUEE Disposable Core Biopsy Instrument, 18g x 20cm, MQ1825 BARD MARQUEE Disposable Core Biopsy Instrument, 18g x 25cm, MQK1210 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 12g x 10cm, MQK1213 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 12g x 13cm, MQK1410 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 14g x 10cm, MQK1413 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 14g x 13cm, MQK1416 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 14g x 16cm, MQK1610 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 16g x 10cm, MQK1810 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 18g x 10cm, MQK1816 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 18g x 16cm, MQK1820 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 18g x 20cm, MQK1825 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 18g x 25cm, MQK2010 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 20g x 10cm, MQK2016 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 20g x 16cm UDI Codes: (01)801741097119(17)230328(10)0001353340, (01)801741097119(17)230328(10)0001353342, (01)801741097119(17)230328(10)0001353401, (01)801741097119(17)230328(10)0001356645, (01)801741097119(17)230328(10)0001358065, (01)801741097119(17)230428(10)0001359983, (01)801741097119(17)230628(10)0001369253, (01)801741097126(17)230328(10)0001353402, (01)801741097126(17)230428(10)0001359984, (01)801741097126(17)230628(10)0001369256, (01)801741097133(17)230328(10)0001353403, (01)801741097133(17)230328(10)0001353404, (01)801741097133(17)230328(10)0001356410, (01)801741097133(17)230328(10)0001356409, (01)801741097133(17)230328(10)0001356886, (01)801741097133(17)230328(10)0001356888, (01)801741097133(17)230428(10)0001359150, (01)801741097133(17)230428(10)0001359152, (01)801741097133(17)230428(10)0001359149, (01)801741097133(17)230428(10)0001359151, (01)801741097133(17)230428(10)0001359988, (01)801741097133(17)230528(10)0001363603, (01)801741097133(17)230528(10)0001363604, (01)801741097133(17)230528(10)0001364078, (01)801741097133(17)230528(10)0001364076, (01)801741097133(17)230528(10)0001364077, (01)801741097133(17)230528(10)0001368474, (01)801741097133(17)230628(10)0001369045, (01)801741097133(17)230628(10)0001369046, (01)801741097140(17)230328(10)0001357081, (01)801741097140(17)230628(10)0001369273, (01)801741097157(17)230328(10)0001357079, (01)801741097164(17)230328(10)0001353246, (01)801741097164(17)230328(10)0001356414, (01)801741097164(17)230428(10)0001359989, (01)801741097164(17)230528(10)0001364079, (01)801741097164(17)230628(10)0001369264, (01)801741097171(17)230328(10)0001356999, (01)801741097171(17)230428(10)0001359990, (01)801741097188(17)230328(10)0001353723, (01)801741097195(17)230328(10)0001353492, (01)801741097195(17)230428(10)0001359991, (01)801741097195(17)230528(10)0001364091, (01)801741097195(17)230628(10)0001369267, (01)801741097201(17)230328(10)0001353722, (01)801741097218(17)230528(10)0001364092, (01)801741097218(17)230528(10)0001364093, (01)801741097218(17)230628(10)0001369270, (01)801741097225(17)230328(10)0001353721, (01)801741097225(17)230328(10)0001355807, (01)801741097225(17)230428(10)0001359994, (01)801741097225(17)230528(10)0001364090, (01)801741097263(17)230328(10)0001353400, (01)801741097263(17)230328(10)0001353399, (01)801741097263(17)230328(10)0001356646, (01)801741097263(17)230428(10)0001359178, (01)801741097263(17)230528(10)0001364069, (01)801741097263(17)230528(10)0001364070, (01)801741097263(17)230528(10)0001366667, (01)801741097263(17)230628(10)0001369252, (01)801741097263(17)230628(10)0001371860, (01)801741097270(17)230528(10)0001364072, (01)801741097270(17)230628(10)0001369259, (01)801741097287(17)230328(10)0001353727, (01)801741097287(17)230328(10)0001356411, (01)801741097287(17)230328(10)0001356412, (01)801741097287(17)230428(10)0001358408, (01)801741097287(17)230428(10)0001358409, (01)801741097287(17)230428(10)0001358414, (01)801741097287(17)230428(10)0001359999, (01)801741097287(17)230428(10)0001358413, (01)801741097287(17)230528(10)0001364084, (01)801741097287(17)230528(10)0001364085, (01)801741097287(17)230528(10)0001364082, (01)801741097287(17)230528(10)0001364080, (01)801741097287(17)230528(10)0001364081, (01)801741097287(17)230528(10)0001367194, (01)801741097287(17)230528(10)0001367195, (01)801741097287(17)230528(10)0001367196, (01)801741097287(17)230528(10)0001368556, (01)801741097287(17)230528(10)0001368555, (01)801741097287(17)230528(10)0001368760, (01)801741097287(17)230628(10)0001369292, (01)801741097294(17)230428(10)0001360001, (01)801741097300(17)230228(10)0001347094, (01)801741097300(17)230328(10)0001357065, (01)801741097317(17)230328(10)0001357859, (01)801741097317(17)230528(10)0001364088, (01)801741097317(17)230628(10)0001369278, (01)801741097348(17)230328(10)0001353725, (01)801741097348(17)230428(10)0001360028, (01)801741097348(17)230628(10)0001369289, (01)801741097355(17)230528(10)0001364089, (01)801741097355(17)230628(10)0001369284, (01)801741097362(17)230628(10)0001369290, (01)801741097379(17)230328(10)0001353724, (01)801741097379(17)230428(10)0001360029, (01)801741097386(17)230628(10)0001369272, (01)801741097393(17)230428(10)0001360030, (01)801741097393(17)230628(10)0001369276, Lot Numbers: 0001353340, 0001353342, 0001353401, 0001356645, 0001358065, 0001359983, 0001369253, 0001353402, 0001359984, 0001369256, 0001353403, 0001353404, 0001356410, 0001356409, 0001356886, 0001356888, 0001359150, 0001359152, 0001359149, 0001359151, 0001359988, 0001363603, 0001363604, 0001364078, 0001364076, 0001364077, 0001368474, 0001369045, 0001369046, 0001357081, 0001369273, 0001357079, 0001353246, 0001356414, 0001359989, 0001364079, 0001369264, 0001356999, 0001359990, 0001353723, 0001353492, 0001359991, 0001364091, 0001369267, 0001353722, 0001364092, 0001364093, 0001369270, 0001353721, 0001355807, 0001359994, 0001364090, 0001353400, 0001353399, 0001356646, 0001359178, 0001364069, 0001364070, 0001366667, 0001369252, 0001371860, 0001364072, 0001369259, 0001353727, 0001356411, 0001356412, 0001358408, 0001358409, 0001358414, 0001359999, 0001358413, 0001364084, 0001364085, 0001364082, 0001364080, 0001364081, 0001367194, 0001367195, 0001367196, 0001368556, 0001368555, 0001368760, 0001369292, 0001360001, 0001347094, 0001357065, 0001357859, 0001364088, 0001369278, 0001353725, 0001360028, 0001369289, 0001364089, 0001369284, 0001369290, 0001353724, 0001360029, 0001369272, 0001360030, 0001369276
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY O.U.S.: Australia, Belgium, Canada, Europe, India, and Singapore
Total 131,876 kits (102,210 kits U.S. and 29,666 kits O.U.S.)