FDA Enforcement Class II Terminated

Helistat Absorbable Hemostatic,Collagen Based Helistat is indicated in surgical procedures (other than ophthalmological and urological surgery) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical.

Recall: Z-1289-2013 · Reported May 22, 2013

Enforcement

Recall Number
Z-1289-2013
Event ID
64778
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Integra LifeSciences Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 22, 2013
Initiation Date
March 28, 2013
Classification Date
May 10, 2013
Termination Date
June 3, 2015
Address
105 Morgan Ln, Plainsboro, NJ, 08536-3339, United States

Description

Helistat Absorbable Hemostatic,Collagen Based Helistat is indicated in surgical procedures (other than ophthalmological and urological surgery) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical.

Reason

Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.

Code Info

1104789 1104790, 1110793 1112248, 1110248 1111867

Distribution

Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.

Quantity

24,342 boxes