LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustable shock absorber and rotation with progressive stiffening -Build height: with male pyramid adapter 145 - 153mm (5 11/16 - 6 ) with female pyramid adapter 153 165mm (6 - 6 1/2) -Male pyramid adapter -Female pyramid adapter -Maintenance free, no lubrication required -Maximum weight: 147kg (325lbs) -Available in sizes 22-30
Enforcement
- Recall Number
- Z-1288-2015
- Event ID
- 70648
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ossur H / F
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 25, 2015
- Initiation Date
- March 4, 2015
- Classification Date
- March 17, 2015
- Termination Date
- October 28, 2015
- Address
- Grjothals 5, Reykjavik, N/A, Iceland
Description
LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustable shock absorber and rotation with progressive stiffening -Build height: with male pyramid adapter 145 - 153mm (5 11/16 - 6 ) with female pyramid adapter 153 165mm (6 - 6 1/2) -Male pyramid adapter -Female pyramid adapter -Maintenance free, no lubrication required -Maximum weight: 147kg (325lbs) -Available in sizes 22-30
The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module can become loose from the foot blade.
LP Rotate, sizes 25-27 cat.1-6. Product numbers LRP062, LRF062, LRPU62, LRFU62. All from manufacturing lot number HF141022
Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,
2,194 (1,464 in U.S. and 730 internationally)