FDA Enforcement
Class II
Ongoing
DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
Recall: Z-1283-2023
·
Reported March 29, 2023
Enforcement
- Recall Number
- Z-1283-2023
- Event ID
- 91760
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- DeRoyal Industries Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 29, 2023
- Initiation Date
- February 9, 2023
- Classification Date
- March 23, 2023
- Address
- 200 Debusk Ln, N/A, Powell, TN, 37849-4703, United States
Description
DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
Reason
The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.
Code Info
GTIN 50749756377046, Lot 58447585 exp. 12/22/2027
Distribution
US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.
Quantity
27,200 devices