AXIOM Artis zee/zeego, system, x-ray, angiographic Product Usage: AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Enforcement
- Recall Number
- Z-1283-2016
- Event ID
- 73672
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 6, 2016
- Initiation Date
- February 18, 2016
- Classification Date
- March 31, 2016
- Termination Date
- August 8, 2017
- Address
- 40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States
Description
AXIOM Artis zee/zeego, system, x-ray, angiographic Product Usage: AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can result in a cooling system failure and subsequent system failure. This leak can occur sporadically in affected systems.
Serial Numbers: Artis Q biplane Serial # 121061, 121089, 121094, 121006, 121013, 121040, 121062, 121082, 121011, 121039, 121134, 121053, 121054, 121014, 121112, 121051, 121004, 121133, 121018, 121036, 121137, 121012, 121049, 121078, 121019, 121010, 121139, 121081, 121096, 121015, 121115, 121016, 121138, 121020, 121140, 121106, 121120, 121122, 121121, 121109, 121144, 121047, 121088, 121027, 121136, 121141, 121069, 121009, 121057, 121056, 121034, 121008, 121058, 121085, 121071, 121023, 121079, 121041, 121037, 121046, 121072
US Nationwide Distribution in the states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY including Puerto Rico.
986 systems in total