FDA Enforcement Class II Terminated

AXIOM Artis zee/zeego, system, x-ray, angiographic Product Usage: AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Recall: Z-1283-2016 · Reported April 6, 2016

Enforcement

Recall Number
Z-1283-2016
Event ID
73672
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 6, 2016
Initiation Date
February 18, 2016
Classification Date
March 31, 2016
Termination Date
August 8, 2017
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

AXIOM Artis zee/zeego, system, x-ray, angiographic Product Usage: AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Reason

Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can result in a cooling system failure and subsequent system failure. This leak can occur sporadically in affected systems.

Code Info

Serial Numbers: Artis Q biplane Serial # 121061, 121089, 121094, 121006, 121013, 121040, 121062, 121082, 121011, 121039, 121134, 121053, 121054, 121014, 121112, 121051, 121004, 121133, 121018, 121036, 121137, 121012, 121049, 121078, 121019, 121010, 121139, 121081, 121096, 121015, 121115, 121016, 121138, 121020, 121140, 121106, 121120, 121122, 121121, 121109, 121144, 121047, 121088, 121027, 121136, 121141, 121069, 121009, 121057, 121056, 121034, 121008, 121058, 121085, 121071, 121023, 121079, 121041, 121037, 121046, 121072

Distribution

US Nationwide Distribution in the states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY including Puerto Rico.

Quantity

986 systems in total