FDA Enforcement Class II Terminated

Siemens LANTIS Oncology Information System Servers; allows the radiation therapist to deliver treatment to patient using the MEVATRON and all available accessories

Recall: Z-1282-2015 · Reported March 25, 2015

Enforcement

Recall Number
Z-1282-2015
Event ID
70384
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 25, 2015
Initiation Date
January 21, 2015
Classification Date
March 16, 2015
Termination Date
July 30, 2015
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406, United States

Description

Siemens LANTIS Oncology Information System Servers; allows the radiation therapist to deliver treatment to patient using the MEVATRON and all available accessories

Reason

There is a potential safety risk when using LANTIS server software with operating systems with which it has not been validated or released which can lead to an incorrect treatment to the patient.

Code Info

Models: 04503178, 05493072, 08148178 with various serial numbers: 93-0938 93-1010 93-0730 93-0454 93-0759 93-0848 93-0884 93-1158 93-0037 93-1053 93-0726 93-0588 93-0693 93-0614 93-0566 93-0648 93-0803 93-0813 93-1387 93-0278 93-0856 93-0657 93-0222 93-1438 93-0662 93-0779 93-0732 93-0841 93-1045 93-0487 93-0564 93-0261 93-0096 93-0846 93-0574 93-0668 93-0887 93-1145 93-0102 93-0764 93-0501 93-0490 93-0739 93-1133 93-0159 93-0872 93-1531 93-0660 93-0798 93-0393 93-0738 93-1296 93-0623 93-0370 93-0951 93-0667 93-1135 93-0141 93-0786 93-0483 93-0385 93-1314 93-0799 93-0456 93-0416 93-0838 93-0639 93-0659 93-0760 93-0608 93-0766 93-0063 93-0878 93-0861 93-0452 93-0533 93-0272 93-0724 93-0596 93-0831 93-0398 93-0870 93-0874 93-0789 93-0686 93-0475 93-1185

Distribution

Nationwide Distribution

Quantity

87