FDA Enforcement Class II Terminated

Dural Graft Matrix DuraGen¿ Dural Graft Matrix is an absorbable implant for repair of dural defects. DuraGen matrix is an easy to handle, soft, white, pliable, nonfriable, porous collagen matrix. DuraGen matrix is supplied sterile, nonpyrogenic, for single use in double peel packages in a variety of sizes..

Recall: Z-1282-2013 · Reported May 22, 2013

Enforcement

Recall Number
Z-1282-2013
Event ID
64778
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Integra LifeSciences Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 22, 2013
Initiation Date
March 28, 2013
Classification Date
May 10, 2013
Termination Date
June 3, 2015
Address
105 Morgan Ln, Plainsboro, NJ, 08536-3339, United States

Description

Dural Graft Matrix DuraGen¿ Dural Graft Matrix is an absorbable implant for repair of dural defects. DuraGen matrix is an easy to handle, soft, white, pliable, nonfriable, porous collagen matrix. DuraGen matrix is supplied sterile, nonpyrogenic, for single use in double peel packages in a variety of sizes..

Reason

Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.

Code Info

1125807, 1125802, 1125806, 1125804, 1125808, 1111845, 1125805, 1125803, 1125809, 1125810, 1130069, 1130061, 1130064, 1130065, 1130417, 1130521, 1110850, 1111839, 1111621, 1110274, 1105302, 1110669, 1111618, 1112228, 1110160, 1110670, 1111606, 1111616, 1104916, 1111614, 1111615, 1110273, 1111277, 1111849, 1110667, 1104918, 1110853, 1111619, 1111842, 1110666, 1112109, 1112108, 1104917, 1110854, 1111620, 1110276, 1105301, 1111846, 1112225, 1110851, 1111617, 1105303, 1110852, 1110270, 1105304, 1110246, 1112222, 1112224, 1110245, 1111844, 1111850, 1110668, 1125522, 1125628, 1104879, 1110814, 1111610, 1125525, 1104876, 1110815, 1111607, 1125526

Distribution

Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.

Quantity

15,607 boxes