FDA Enforcement Class II Terminated

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Recall: Z-1281-2019 · Reported May 29, 2019

Enforcement

Recall Number
Z-1281-2019
Event ID
82753
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Neusoft Medical Systems Co., Ltd.
Voluntary / Mandated
FDA Mandated
Initial Notification
Letter
Report Date
May 29, 2019
Initiation Date
October 22, 2018
Classification Date
May 21, 2019
Termination Date
August 25, 2022
Address
No. 177-1 Cuangxin Road, Hunnan District, Shenyang, N/A, N/A, China

Description

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Reason

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Code Info

Series numbers N16E180037E; N16E180047E

Distribution

US Nationwide Distribution.

Quantity

2