FDA Enforcement Class II Terminated

Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.

Recall: Z-1281-2015 · Reported March 25, 2015

Enforcement

Recall Number
Z-1281-2015
Event ID
70455
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 25, 2015
Initiation Date
February 17, 2015
Classification Date
March 16, 2015
Termination Date
August 26, 2016
Address
1302 Wrights Ln E, West Chester, PA, 19380-3417, United States

Description

Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.

Reason

For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.

Code Info

Part numbers:03.010.048 and 03.010.052 with Lot Numbers: 4874015 4874021 4927262 4954481 4963192 4874016 4874022 4944750 4957586 4963193 4784017 4899564 4944751 4957587 4972003 4874018 4899565 4954478 4957588 4972004 4874019 4927260 4954479 4957589 4874020 4927261 4954480 4963190 4922533 4922975 4981158 4922970 4923015 4981159 4922971 4977241 4981160 4922972 4977242 4981162 4922973 4981156 5035905 4922974 4981157 5035906

Distribution

Nationwide Distribution

Quantity

183