FDA Enforcement Class II Terminated

CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Enteral Feeding tube, ANTI-IV* Connector, 10 Fr, 140 cm The device is used in patients that require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.

Recall: Z-1277-2018 · Reported April 11, 2018

Enforcement

Recall Number
Z-1277-2018
Event ID
79479
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Halyard Health, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 11, 2018
Initiation Date
December 13, 2017
Classification Date
March 30, 2018
Termination Date
November 4, 2021
Address
5405 Windward Pkwy, N/A, Alpharetta, GA, 30004-3894, United States

Description

CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Enteral Feeding tube, ANTI-IV* Connector, 10 Fr, 140 cm The device is used in patients that require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.

Reason

The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube. Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window.

Code Info

Product code 20-9551TRAK2A, Lot No. 7909, 7839, 7876.

Distribution

Worldwide Distribution: US (nationwide) to states of: AR, CA, CO, GA, IA, LA, MI, MN, NM, TX; and country of Germany. Distributed internationally to Germany.

Quantity

185 cases (10 units per case)