FDA Enforcement Class II Terminated

Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA Assays. This recall is for any Phadia Prime software version up to and including the current version, 2.1.4.

Recall: Z-1276-2018 · Reported April 11, 2018

Enforcement

Recall Number
Z-1276-2018
Event ID
79438
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Phadia Ab
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 11, 2018
Initiation Date
November 20, 2017
Classification Date
March 30, 2018
Termination Date
March 6, 2019
Address
Rapsgatan Plant, Rapsgatan 7, Uppsala, N/A, N/A, Sweden

Description

Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA Assays. This recall is for any Phadia Prime software version up to and including the current version, 2.1.4.

Reason

We want to inform all Phadia 250 system operators performing EliA Assays not to use the function OK to All in any version of Phadia Prime, up to and including 2.1.4, when rejecting and retesting samples with any EliA assay (the OK function may be used for rejecting single tests and dilution of samples in accordance with product DfU).

Code Info

Phadia Prime software 12-4101-00 Phadia 250 12-3900-01 Listed software: EliA dsDNA Well 14-5500-01 EliA U1RNP Well 14-5501-01 EliA Sm Well 14-5502-01 EliA Ro Well 14-5503-01 EliA La Well 14-5504-01 EliA CENP Well 14-5505-01 EliA Scl-70 Well 14-5506-01 EliA Jo-1 Well 14-5507-01 EliA Symphony Well 14-5508-01 EliA RNP70 Well 14-5511-01 EliA GBM Well 14-5514-01 EliA CCP Well 14-5515-01 EliA Celikey IgA Well 14-5517-01 EliA Celikey IgG Well 14-5518-01 EliA Gliadin IgA Well 14-5519-01 EliA Gliadin IgG Well 14-5520-01 EliA Cardiolipin IgA Well 14-5528-01 EliA Cardiolipin IgG Well 14-5529-01 EliA Cardiolipin IgM Well 14-5530-01 EliA B2-Glycoprotein I IgA Well 14-5531-01 EliA B2-Glycoprotein I IgG Well 14-5532-01 EliA B2-Glycoprotein I IgM Well 14-5533-01 EliA PR3S Well 14-5536-01 EliA MPOS Well 14-5537-01 EliA GliadinDP IgA Well 14-5538-01 EliA GliadinDP IgG Well 14-5539-01 EliA RF IgM Well 14-5600-01 EliA RF IgA Well 14-5601-01 EliA SmDP Well 14-5624-01 EliA Scl-70S Well 14-5637-01 EliA anti-TPO Well 14-5641-01 EliA anti-TG Well 14-5642-01 EliA M2 Well 14-5649-01

Distribution

US Distribution to the states of :TX, VA, NJ, MA, IN, GA, UT, TN, CA, MS, NC, ME and OR

Quantity

2307