TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.
Enforcement
- Recall Number
- Z-1273-2020
- Event ID
- 84748
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Smith & Nephew, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 26, 2020
- Initiation Date
- December 30, 2019
- Classification Date
- February 19, 2020
- Termination Date
- May 11, 2021
- Address
- 130 & 120 Forbes Blvd, Mansfield, MA, 02048, United States
Description
TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.
Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.
Batch numbers: 2010547, 2025137, 2036695, 50763859, & 50773071. Part Number 72200796. UDI 10600147.
Worldwide distribution US Nationwide distribution in the states of AZ, ID, MO, NC, ND, NE, OK, SC, TN, and countries of AE, AU, CA, CH, ES, FI, JP, PR, SG.
82 distributed throughout US:AZ, ID, MO, NC, ND, NE, OK, SC, TN. 243 distributed OUS: AE, AU, CA, CH, ES, FI, JP, PR, SG. Total: 325 units