FDA Enforcement Class II Terminated

Ellex Medical Fixed eye Safety filters Model Integre S LP5532. The filter is located in the binocular mount of the laser delivery head. The filter prevents transmission of treatment laser light to physician eyes while the aiming beam to remain visible. .

Recall: Z-1270-2016 · Reported April 13, 2016

Enforcement

Recall Number
Z-1270-2016
Event ID
73468
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Laserex Systems Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 13, 2016
Initiation Date
February 1, 2016
Classification Date
April 7, 2016
Termination Date
October 27, 2016
Address
7138 SHADY OAK RD, N/A, EDEN PRAIRIE, MN, 55344-3517, United States

Description

Ellex Medical Fixed eye Safety filters Model Integre S LP5532. The filter is located in the binocular mount of the laser delivery head. The filter prevents transmission of treatment laser light to physician eyes while the aiming beam to remain visible. .

Reason

The Firm discovered that there is a design error which is traceable back to 2012 when a design change on the safety filter optic adaptor was implemented.

Code Info

Model Number : 3143270

Distribution

Worldwide Distribution - US Distribution

Quantity

72 units plus 2 spare units