Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer
Enforcement
- Recall Number
- Z-1270-2014
- Event ID
- 67607
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Lin-Zhi International Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 2, 2014
- Initiation Date
- February 18, 2014
- Classification Date
- March 21, 2014
- Termination Date
- April 25, 2014
- Address
- 670 Almanor Ave, N/A, Sunnyvale, CA, 94085-3513, United States
Description
Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer
Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).
Product catalog number: A68827; Lots 1308006, 1308131, Expiry: March 28, 2015.
US Distribution including CA.
120 units, 60 of each lot number