FDA Enforcement
Class II
Ongoing
CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)
Recall: Z-1269-2025
·
Reported March 12, 2025
Enforcement
- Recall Number
- Z-1269-2025
- Event ID
- 96224
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Kung Shin Plastics Co. Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 12, 2025
- Initiation Date
- January 9, 2025
- Classification Date
- March 4, 2025
- Address
- 65-2 Erh Chia Road, Ying Kuo Town, Taipei Hsien, N/A, Taiwan
Description
CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)
Reason
The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.
Code Info
UDI (01)00810050853682(10)C2N24002(11)240828(17)290827 Lot C2N24002
Distribution
US Nationwide distribution in the state of IL.
Quantity
30000 ea/ 300 cases