FDA Enforcement Class II Terminated

MicroScan LabPro Information Manager System, Software Versions 1.0 to 4.41 Catalog No. 10638819, 10638820, 10638823, 10638824, 10638825,10638826, 10714149,10714159, 10805072, 10805073, 10975000, 10975001, 6000-0008, 6000-0026. Japan: 10638819 LabPro v4.11 Software Update Kit 10638820 LabPro v4.11 System Software 10805072 LabPro MBT Non-Connect Systems v4.30 10805073 LabPro MBT Connect Systems v4.30 Canada: 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit Europe: 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software USA: 10714149, LabPro v4.11 Software Update Kit 10714150 LabPro v4.11 System Software 10975000 ASM LabPro-MBT v4.40 10975001 ASM LabPro-MBT v4.40 for Connect 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit ROW: 6000-0008 LabPro v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software Product Usage: LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel. LabPro AlertEx is a functional subset of the LabPro Data Management System that analyzes MicroScan ID and AST data, or other predefined parameters, against a series of established rules/alerts and notifies the user of unusual and/or critical conditions, which may warrant further analysis or actions.

Recall: Z-1269-2016 · Reported April 6, 2016

Enforcement

Recall Number
Z-1269-2016
Event ID
73417
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 6, 2016
Initiation Date
February 11, 2016
Classification Date
March 28, 2016
Termination Date
January 4, 2018
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

MicroScan LabPro Information Manager System, Software Versions 1.0 to 4.41 Catalog No. 10638819, 10638820, 10638823, 10638824, 10638825,10638826, 10714149,10714159, 10805072, 10805073, 10975000, 10975001, 6000-0008, 6000-0026. Japan: 10638819 LabPro v4.11 Software Update Kit 10638820 LabPro v4.11 System Software 10805072 LabPro MBT Non-Connect Systems v4.30 10805073 LabPro MBT Connect Systems v4.30 Canada: 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit Europe: 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software USA: 10714149, LabPro v4.11 Software Update Kit 10714150 LabPro v4.11 System Software 10975000 ASM LabPro-MBT v4.40 10975001 ASM LabPro-MBT v4.40 for Connect 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit ROW: 6000-0008 LabPro v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software Product Usage: LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel. LabPro AlertEx is a functional subset of the LabPro Data Management System that analyzes MicroScan ID and AST data, or other predefined parameters, against a series of established rules/alerts and notifies the user of unusual and/or critical conditions, which may warrant further analysis or actions.

Reason

Beckman Coulter is recalling the MicroScan LabPro Information Manager System because the software incorrectly allows the operator to manually edit the carbohydrate substrates when manually reading dried overnight gram negative panels with an ID Hold status.

Code Info

not applicable

Distribution

Affected geographies: Bahamas, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Lithuania, Mexico, Morocco, Netherlands, Northern Mariana Islands, Panama, Paraguay, Peru, Poland, Portugal, Puerto Rico, Romania, Russia, Singapore, South Africa, Spain, Taiwan, Tunisia, Turkey, United States, Uruguay, Virgin Islands, Zimbabwe.

Quantity

2,702 units total (1,032 units in US)