FDA Enforcement Class II Terminated

VITROS SARS-CoV-2 Antigen Reagent Pack - The VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack when used in combination with the VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) specimens. For emergency authorization use only (EAU). Product Code: 6199941. OUS Product.

Recall: Z-1267-2021 · Reported March 24, 2021

Enforcement

Recall Number
Z-1267-2021
Event ID
87356
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 24, 2021
Initiation Date
February 12, 2021
Classification Date
March 17, 2021
Termination Date
January 24, 2022
Address
100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States

Description

VITROS SARS-CoV-2 Antigen Reagent Pack - The VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack when used in combination with the VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) specimens. For emergency authorization use only (EAU). Product Code: 6199941. OUS Product.

Reason

VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results Associated When Using Remel M4RT Viral Transport Media

Code Info

Any lot used with Remel M4RT VTM UDI: 10758750033584

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Canada L3R 4G5, Chile, Colombia, Denmark, France, Germany, India, Italy, Japan, Portugal, Russia, Singapore 609917, Spain, United Kingdom.

Quantity

12177 units