FDA Enforcement
Class II
Terminated
EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.
Recall: Z-1267-2018
·
Reported April 4, 2018
Enforcement
- Recall Number
- Z-1267-2018
- Event ID
- 77160
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medica Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 4, 2018
- Initiation Date
- April 11, 2017
- Classification Date
- March 28, 2018
- Termination Date
- August 20, 2019
- Address
- 5 Oak Park Dr, Bedford, MA, 01730-1430, United States
Description
EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.
Reason
Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.
Code Info
Medica s Lot # 16292
Distribution
USA: MI FL CA TX India, Russia, Puerto Rico, N. Mariana Island/Palau.