FDA Enforcement Class II Terminated

Basic Pack, part number AMS1868(B Basic Pack, part number AMS2608(A Basic Pack, part number AMS3861 Basic Pack, part number AMS4309 Basic Pack, part number PSS1833 Basic Pack, part number PSS1833(A Basic Pack, part number PSS1848(A

Recall: Z-1267-2017 · Reported March 8, 2017

Enforcement

Recall Number
Z-1267-2017
Event ID
76280
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Windstone Medical Packaging, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
March 8, 2017
Initiation Date
January 11, 2017
Classification Date
February 28, 2017
Termination Date
May 3, 2017
Address
1602 4th Ave N, N/A, Billings, MT, 59101-1521, United States

Description

Basic Pack, part number AMS1868(B Basic Pack, part number AMS2608(A Basic Pack, part number AMS3861 Basic Pack, part number AMS4309 Basic Pack, part number PSS1833 Basic Pack, part number PSS1833(A Basic Pack, part number PSS1848(A

Reason

The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES of the custom Aligned Medical Solutions (AMS) kits were recalled because gloves may contain splits or holes compromising sterility.

Code Info

Lot number and expiration date 100394, 9/15/2019 100593, 9/21/2019 100594, 10/5/2019 101063, 10/15/2019 101304, 10/23/2019 83725, 6/8/2018 84749, 9/11/2018 85159, 8/1/2018 85635, 8/6/2018 85810, 7/19/2018 86252, 9/2/2018 86440, 8/7/2018 86859, 8/18/2018 86939, 11/22/2018 87105, 9/15/2018 87269, 9/30/2018 87351, 9/15/2018 87447, 11/11/2018 87942, 9/23/2018 88172, 12/3/2018 88282, 9/7/2018 88368, 9/5/2018 88846, 12/11/2018 88887, 12/7/2018 89004, 10/8/2018 89160, 12/1/2018 89438, 11/22/2018 89512, 10/9/2018 89716, 1/9/2019 89743, 12/3/2018 89811, 12/1/2018 94572, 12/15/2018 94705, 12/15/2018 94751, 1/4/2019 94853, 1/27/2019 94935, 1/25/2019 95293, 1/26/2019 95412, 1/25/2019 95936, 1/11/2019 95974, 5/29/2019 96023, 1/16/2019 96255, 1/2/2019 96410, 3/7/2019 96566, 3/9/2019 96678, 5/24/2019 96913, 5/22/2019 97124, 5/12/2019 97410, 6/3/2019 97577, 5/1/2019 97780, 6/15/2019 98017, 6/29/2019 98312, 6/24/2019 98478, 6/29/2019 98502, 6/6/2019 98663, 6/28/2019 98633, 6/30/2019 99172, 6/12/2019 99581, 6/27/2019 99703, 9/27/2019 99843, 9/30/2019 99887, 9/11/2019 99913, 9/23/2019

Distribution

Distributed in AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NJ, NV, OH, PA, TX, UT, VA, and WA.

Quantity

2388 kits