FDA Enforcement Class II Ongoing

Attune Revision Limb Preservation System (LPS) Tibial Insert XSM 26MM, Part Number 151760226

Recall: Z-1265-2023 · Reported March 29, 2023

Enforcement

Recall Number
Z-1265-2023
Event ID
91752
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
DePuy Orthopaedics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 29, 2023
Initiation Date
February 15, 2023
Classification Date
March 17, 2023
Address
700 Orthopaedic Dr, N/A, Warsaw, IN, 46582-3994, United States

Description

Attune Revision Limb Preservation System (LPS) Tibial Insert XSM 26MM, Part Number 151760226

Reason

The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.

Code Info

GTIN: 10603295491033; Lot: JP9022

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Poland, Germany, India, Australia, & New Zealand

Quantity

7 units