i-STAT G3+ cartridge (blue), List No. 03P78-50, UDI (01) 00054749002283 - Product Usage: The i-STAT G3+ cartridge may be used for the quantitative measurement of pH, pCO2, and pO2 in venous, arterial or capillary whole blood.
Recall: Z-1265-2020
·
Reported February 26, 2020
Enforcement
- Recall Number
- Z-1265-2020
- Event ID
- 84730
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Point Of Care Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- February 26, 2020
- Initiation Date
- January 14, 2020
- Classification Date
- February 18, 2020
- Termination Date
- April 12, 2021
- Address
- 400 College Rd E, Princeton, NJ, 08540-6607, United States
Description
i-STAT G3+ cartridge (blue), List No. 03P78-50, UDI (01) 00054749002283 - Product Usage: The i-STAT G3+ cartridge may be used for the quantitative measurement of pH, pCO2, and pO2 in venous, arterial or capillary whole blood.
Reason
Marketed without a 510(k). Abbott Point of Care has determined that it does not have sufficient clinical evidence to fully characterize the performance of the blue i-STAT G3+ cartridges and therefore will no longer distribute the i-STAT G3+ (blue) cartridge in the United States as of May 1, 2020.
Code Info
All lots are affected
Distribution
US Nationwide distribution.
Quantity
4,448,525