FDA Enforcement
Class II
Ongoing
Attune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 151760212
Recall: Z-1264-2023
·
Reported March 29, 2023
Enforcement
- Recall Number
- Z-1264-2023
- Event ID
- 91752
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- DePuy Orthopaedics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 29, 2023
- Initiation Date
- February 15, 2023
- Classification Date
- March 17, 2023
- Address
- 700 Orthopaedic Dr, N/A, Warsaw, IN, 46582-3994, United States
Description
Attune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 151760212
Reason
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.
Code Info
GTIN: 10603295490968; Lot: JP9016
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Poland, Germany, India, Australia, & New Zealand
Quantity
7 units