FDA Enforcement Class II Terminated

NM/CT 860 Product Usage: The GE NM/CT 860 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body and vascular X-ray Computed Tomography applications.

Recall: Z-1263-2019 · Reported May 8, 2019

Enforcement

Recall Number
Z-1263-2019
Event ID
82582
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
GE Healthcare, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 8, 2019
Initiation Date
March 20, 2019
Classification Date
May 2, 2019
Termination Date
October 5, 2020
Address
3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States

Description

NM/CT 860 Product Usage: The GE NM/CT 860 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body and vascular X-ray Computed Tomography applications.

Reason

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Code Info

Mfg. Lot or Serial # System ID Model, Catalogue or Code # 860Y62005 SHIPPED (GON 4736850) H3908AD 850X61003 401793860NM H3908AD 860X62001 E6843969 H3908AE 870X64005 HC4387NU05 H3100AS 860Y62004 083026212959219 H3908AD 860X62002 PL0399NM01 H3908AD

Distribution

US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK

Quantity

6 units