ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator - product Usage: For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the CardioPhase¿ hsCRP assay.
Enforcement
- Recall Number
- Z-1260-2020
- Event ID
- 84413
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 26, 2020
- Initiation Date
- November 19, 2019
- Classification Date
- February 18, 2020
- Termination Date
- August 30, 2021
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005, United States
Description
ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator - product Usage: For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the CardioPhase¿ hsCRP assay.
Atellica and ADVIA hsCRP calibrator accuracy has drifted away from the reference materials (ERM-DA470/IFCC), which is causing a high bias on QC and patient sample recovery.
Siemens Material Number (SMN): ADVIA Chemistry: 10335897 REF: 05006455 Lot# 453610 Expiration Date: 2020-01 UDI Number: ADVIA: (01)00630414526843(10)453610(17)20200128
Worldwide distribution US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WY and countries of UK, CA, PH, MX, VN, TW, TH, IN, CO, AE, AR, AU, BD, BE, BG, BR, CL, CZ, DE, DK, EG, ES, FI, FR, GR, HU, IE, IT, KR, MY, NL, PL, PT, RO, RU, SE, SG.