FDA Enforcement
Class II
Ongoing
Brainlab ExacTrac Dynamic software, Medical Charged-Particle Radiation Therapy System, model 20910-01E and 20910-01F.
Recall: Z-1259-2023
·
Reported March 22, 2023
Enforcement
- Recall Number
- Z-1259-2023
- Event ID
- 91664
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Brainlab AG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 22, 2023
- Initiation Date
- February 14, 2023
- Classification Date
- March 16, 2023
- Address
- Olof-palme-str. 9, N/A, Munich, N/A, N/A, Germany
Description
Brainlab ExacTrac Dynamic software, Medical Charged-Particle Radiation Therapy System, model 20910-01E and 20910-01F.
Reason
Deep Inspiration Breath Hold (DIBH) functionality may not work as specified when based on a standard-workflow Treatment Template
Code Info
Versions 1.1.0 and 1.1.1, GTIN: 04056481143978.
Distribution
US Nationwide distribution in the states of AZ, AR, CA, CO, FL, GA. IL, IN, KY, LA, MS, MO, NJ, NY, NC, ND, OK, OR, PA, TN, TX, VA, and WA. There was no government/military distribution.
Quantity
56 systems in the U.S.