FDA Enforcement
Class II
Ongoing
t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology
Recall: Z-1258-2022
·
Reported June 22, 2022
Enforcement
- Recall Number
- Z-1258-2022
- Event ID
- 90113
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Tandem Diabetes Care Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 22, 2022
- Initiation Date
- September 10, 2020
- Classification Date
- June 10, 2022
- Address
- 11075 Roselle St, San Diego, CA, 92121-1204, United States
Description
t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology
Reason
Insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to less than 5%, insulin delivery will continue for 30 minutes and then the pump will power off and insulin delivery will stop. If the battery reaches 1%, then insulin delivery will stop, which could lead to hyperglycemia.
Code Info
Dexcom G5 - All Serial Numbers; Basal-IQ Technology - Software Versions 6.3.01 and Earlier
Distribution
International distribution in the country of Australia.
Quantity
5,206