Natural-Knee¿ II Posterior Stabilized (P.S.) Condylar Tibial Inserts: N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS INS TIB SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ
Enforcement
- Recall Number
- Z-1257-2014
- Event ID
- 67503
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- March 26, 2014
- Initiation Date
- February 25, 2014
- Classification Date
- March 20, 2014
- Termination Date
- March 24, 2015
- Address
- 1800 W Center St, Warsaw, IN, 46580-2304, United States
Description
Natural-Knee¿ II Posterior Stabilized (P.S.) Condylar Tibial Inserts: N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS INS TIB SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Item Number 672509100 672509101 672509200 672509203 672511100 672511103 672513101 672513103 672513201 672516103 672519103 672522201 672525103 672525203
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
3638 units