FDA Enforcement
Class III
Terminated
Ethyl Alcohol Enzymatic Assay, (a) REF 0220 (small test kit), (b) REF 0221 (large test kit) Product Usage: Detergent causes faster degradation in EtOH assay. Possible container or pitchers with some detergent residue was used while producing R2 of the defective lot.
Recall: Z-1256-2019
·
Reported May 8, 2019
Enforcement
- Recall Number
- Z-1256-2019
- Event ID
- 82011
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Lin-Zhi International Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- May 8, 2019
- Initiation Date
- January 11, 2019
- Classification Date
- May 2, 2019
- Termination Date
- March 22, 2021
- Address
- 2945 Oakmead Village Ct, N/A, Santa Clara, CA, 95051-0812, United States
Description
Ethyl Alcohol Enzymatic Assay, (a) REF 0220 (small test kit), (b) REF 0221 (large test kit) Product Usage: Detergent causes faster degradation in EtOH assay. Possible container or pitchers with some detergent residue was used while producing R2 of the defective lot.
Reason
Shelf life of the product may be reduced due to degradation of the assay.
Code Info
(a) REF 0220 (small test kit), Lot Code 1810026 (b) REF 0221 (large test kit), Lot Code 1810025
Distribution
US Nationwide Distribution in the states of CA, MA, GA, KY, NC, MI, MD, NC, CT, AZ SC, TN, OR
Quantity
49 kits