Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.
Enforcement
- Recall Number
- Z-1255-2021
- Event ID
- 87223
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- R & D Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 24, 2021
- Initiation Date
- January 20, 2021
- Classification Date
- March 15, 2021
- Termination Date
- February 13, 2023
- Address
- 614 Mckinley Pl Ne, N/A, Minneapolis, MN, 55413-2610, United States
Description
Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.
R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.
Catalog #: DTFR1; Unique Device Identification (UDI) #: 815762021657; Lot #: P252478; P263156; P256349; P263156; P263160; P264604
USA: (FL, LA, CA, OH), CHINA, UNITED KINGDOM, JAPAN, SINGAPORE, Jakarta Selatan,
156 kits