RAPIDLab 1265 System; Siemens Material Number: 10491395, 10321852, 10335524. The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb.
Enforcement
- Recall Number
- Z-1255-2015
- Event ID
- 70203
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 18, 2015
- Initiation Date
- December 30, 2014
- Classification Date
- March 6, 2015
- Termination Date
- August 2, 2017
- Address
- 2 Edgewater Drive, N/A, Norwood, MA, 02062, United States
Description
RAPIDLab 1265 System; Siemens Material Number: 10491395, 10321852, 10335524. The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb.
Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
All serial numbers.
Worldwide Distribution - US nationwide including Puerto Rico, Canada, Mexico, Albania, Algeria, Argentina, Austria, Bahrain, Belgium, Bosnia Herzegovina, Botswana, Brazil, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iraq, Ireland, Israel, Italy, Ivory Coast, Kazakhstan, Kuwait, Latvia, Lebanon, Lesotho, Lithuania, Luxembourg, Macedonia, Netherlands, New Caledonia, Norway, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, U.A.E., United Kingdom, Vatican City, Venezuela, White Russia/Belarus.
2863 devices