FDA Enforcement Class II Terminated

Natural-Knee¿ Congruent Tibial Insert with Screw: SF-CON TIB INS LT SZ1/2 1 SF-CON TIB INS RT SZ1/2 9 SF-CON TIB INS RT SZ3/4/5

Recall: Z-1254-2014 · Reported March 26, 2014

Enforcement

Recall Number
Z-1254-2014
Event ID
67503
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
March 26, 2014
Initiation Date
February 25, 2014
Classification Date
March 20, 2014
Termination Date
March 24, 2015
Address
1800 W Center St, Warsaw, IN, 46580-2304, United States

Description

Natural-Knee¿ Congruent Tibial Insert with Screw: SF-CON TIB INS LT SZ1/2 1 SF-CON TIB INS RT SZ1/2 9 SF-CON TIB INS RT SZ3/4/5

Reason

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Code Info

Item Number 620005819 620005909 620006916

Distribution

Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.

Quantity

3638 units