FDA Enforcement Class II Ongoing

Brand Name: TBS Osteo Product Name: TBS iNsight Model/Catalog Number: TBS iNsight v3.x / pn: OS-TBSi Software Version: V3.x: (1) V3.1.2, (2) V3.1.1, (3) V3.1.0, (4)V3.0.0 Product Description: TBS iNsight is a class II Medical Device software that is installed on bone densitometers for analysis of bone microarchitecture and bone health management Component: No

Recall: Z-1250-2025 · Reported March 12, 2025

Enforcement

Recall Number
Z-1250-2025
Event ID
96233
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medimaps Group
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 12, 2025
Initiation Date
February 3, 2025
Classification Date
February 28, 2025
Address
Fongit, Chemin des Aulx 18, Plan-les-Ouates, N/A, Switzerland

Description

Brand Name: TBS Osteo Product Name: TBS iNsight Model/Catalog Number: TBS iNsight v3.x / pn: OS-TBSi Software Version: V3.x: (1) V3.1.2, (2) V3.1.1, (3) V3.1.0, (4)V3.0.0 Product Description: TBS iNsight is a class II Medical Device software that is installed on bone densitometers for analysis of bone microarchitecture and bone health management Component: No

Reason

Potential variability in calculations from fast array scans compared to array scans when operating on Hologic Horizon machines.

Code Info

UDI-DI: (1) B214OSTBSI3120, (2) B214OSTBSI3110, (3) B214OSTBSIV310, (4) B214OSTBSIV300.

Distribution

Domestic: AL, AZ, CA, CO, DE, FL, IL, LA, MA, MD, MI, MN, MO, MS, NC, NM, NY, OH, PA, SC, TX, UT, WA. International: ALGERIA, ANDORRA, AUSTRALIA, AUSTRIA, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, CHILE, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, SOUTH KOREA, LEBANON, MALAYSIA, MEXICO, MONTENEGRO, MOROCCO, NEPAL, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, ROMANIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY

Quantity

2036 units (724 US, 1312 OUS)