MRI system: Vantage Elan, Model: MRT-2020
Recall: Z-1249-2024
·
Reported March 13, 2024
Enforcement
- Recall Number
- Z-1249-2024
- Event ID
- 93874
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Canon Medical System, USA, INC.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 13, 2024
- Initiation Date
- January 29, 2024
- Classification Date
- March 4, 2024
- Address
- 2441 Michelle Dr, Tustin, CA, 92780-7047, United States
Description
MRI system: Vantage Elan, Model: MRT-2020
Reason
For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
Code Info
UDI-DI: 04987670101632. Serial Number: S1F2052753
Distribution
Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.
Quantity
1