Gender Solutions" Natural-Knee¿ Flex System: N-K FLEX MBACK SZ 0 PAT N-K FLEX MBACK SZ 1 PAT N-K FLEX MBACK SZ 2 PAT
Enforcement
- Recall Number
- Z-1248-2014
- Event ID
- 67503
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- March 26, 2014
- Initiation Date
- February 25, 2014
- Classification Date
- March 20, 2014
- Termination Date
- March 24, 2015
- Address
- 1800 W Center St, Warsaw, IN, 46580-2304, United States
Description
Gender Solutions" Natural-Knee¿ Flex System: N-K FLEX MBACK SZ 0 PAT N-K FLEX MBACK SZ 1 PAT N-K FLEX MBACK SZ 2 PAT
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Item Numbers 00541800000 00541800001 00541800002
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
3638 units