FDA Enforcement Class II Terminated

2.5MM Aggressive Blade Known as a Sinuscopic Shaver Blade, this device is composed of a metal cutter and plastic drive shaft that goes into a metal housing supported by a plastic hub. ENT Precision Cutter, Aggressive, ESSxlHUMMER 4, Rx Only It is used for ENT procedures and functions by means of rotation when adapted to a handpiece and console. Through the rotation, the necessary tissue is resected as required by the user. The blade housing has an etching line guide that starts with a straight line and along with a scale from 1cm to 7cm in 1cm increments that serves the doctor as a guide by indicating how far the device is being introduced into the surgical area.

Recall: Z-1247-2014 · Reported March 26, 2014

Enforcement

Recall Number
Z-1247-2014
Event ID
67775
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Instruments Div. of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 26, 2014
Initiation Date
March 18, 2014
Classification Date
March 20, 2014
Termination Date
May 15, 2014
Address
4100 E Milham Ave, Portage, MI, 49002-9704, United States

Description

2.5MM Aggressive Blade Known as a Sinuscopic Shaver Blade, this device is composed of a metal cutter and plastic drive shaft that goes into a metal housing supported by a plastic hub. ENT Precision Cutter, Aggressive, ESSxlHUMMER 4, Rx Only It is used for ENT procedures and functions by means of rotation when adapted to a handpiece and console. Through the rotation, the necessary tissue is resected as required by the user. The blade housing has an etching line guide that starts with a straight line and along with a scale from 1cm to 7cm in 1cm increments that serves the doctor as a guide by indicating how far the device is being introduced into the surgical area.

Reason

Stryker Instruments is voluntarily recalling the single use 2.5MM Aggressive Blade because the line on the affected part (P/N 5290-928-000, Lot #13280CG2) was a straight line without a number scale or logo. The line required for this part number includes a number scale that goes from 1cm  7cm.

Code Info

P/N 5290-628-000, Lot #13280CG2

Distribution

US Distribution including the states of NV, CA, WA, TN, OR, SD, ID, IN and CO., and Internationally to Malaysia

Quantity

90 units