FDA Enforcement Class I Ongoing

Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20, PED3-021-250-12, PED3-021-300-14, PED3-021-250-16, PED3-021-300-12, PED3-021-275-14, PED3-021-350-14, PED3-021-350-25, PED3-021-350-12, PED3-021-250-10, PED3-021-275-20, PED3-021-350-18, PED3-021-325-18, PED3-021-275-18, PED3-021-250-18, PED3-021-300-18; and Instructions for Use, Part: M056989CDOC2

Recall: Z-1246-2025 · Reported March 12, 2025

Enforcement

Recall Number
Z-1246-2025
Event ID
96168
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Micro Therapeutics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 12, 2025
Initiation Date
January 30, 2025
Classification Date
March 5, 2025
Address
9775 Toledo Way, Irvine, CA, 92618-1811, United States

Description

Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20, PED3-021-250-12, PED3-021-300-14, PED3-021-250-16, PED3-021-300-12, PED3-021-275-14, PED3-021-350-14, PED3-021-350-25, PED3-021-350-12, PED3-021-250-10, PED3-021-275-20, PED3-021-350-18, PED3-021-325-18, PED3-021-275-18, PED3-021-250-18, PED3-021-300-18; and Instructions for Use, Part: M056989CDOC2

Reason

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age

Code Info

Instructions for Use, Version: Rev. A (01/2024). REF/UDI-DI/Lot: PED3-021-350-16/00847536030701, 00847536035195, 00847536036529/B317460, B327449, B335184, B340948, B349032, B389162, B392918, B424454, B430015, B495821, B497938, B547396, B550307, B551615, B559033, B603377, B623564, B656863, B656887, B658904, B692346, B715057, B715058, B715541, B726894, B735821, B735823, B756029, B756030, B760210, B760211, B768202, B768387, B769453, B770633, B807494; PED3-021-250-14/00847536030411, 00847536034907, 00847536036239/B317936 B332767, B333492, B333536, B334545, B376967, B388041, B391185, B399354, B416713, B431395, B452416, B457963, B467691, B476239, B564173, B564970, B574472, B580923, B629372, B630998, B670305, B673035, B699576, B707568, B709193, B736698, B736893, B739090, B739091, B795110, B796586, B800827, B801770; PED3-021-350-20/00847536030725, 00847536035218, 00847536036543/B319436, B336571, B379993, B383427, B389708, B400103, B400599, B403612, B407062, B415211, B446722, B483987, B490972, B492403, B494049, B552487, B559741, B591782, B591864, B592074, B622656, B626023, B648708, B684905, B715692, B719565, B719567, B719659, B726923, B728926, B735667, B756330, B765053, B768231, B770237, B770248, B792630; PED3-021-250-20/00847536030442, 00847536034938, 00847536036260/B321465, B337378, B360749, B371070, B412868, B467843, B477487, B488713, B548244, B557602, B644034, B673388, B679822, B684208, B686727, B695486, B698571, B703035, B708080, B709204, B711078, B746067, B757056, B757060, B761030, B771498, B775215, B789892, B790055; PED3-021-300-16/00847536030541, 00847536035034, 00847536036369/B321725, B338497, B405720, B418200, B570723, B574473, B603375, B604477, B656277, B658580, B659806, B666425, B681289, B691752, B699606, B703209, B707982, B711365, B725679, B736659, B738114, B769551, B769582, B798864; PED3-021-300-20/00847536030565, 00847536035058, 00847536036383/B322263, B336066, B347327, B376257, B405892, B414535, B458672, B458776, B469040, B490920, B495292, B563197, B607500, B607501, B635214, B636992, B656299, B656355, B657960, B668529, B678482, B680045, B681190, B712110, B712111, B718522, B720923, B725457, B732895, B735956, B737156, B737367, B748272, B748683, B749541, B775230, B778774, B780044; PED3-021-275-12/00847536030466, 00847536034952, 00847536036284/B323044, B383016, B391275, B399486, B454385, B468406, B623580, B623583, B624424, B657776, B666498, B669520, B670179, B670226, B710976, B743820, B743822, B754659, B762034, B768199, B798851; PED3-021-275-16/00847536030480, 00847536034976, 00847536036307/B323046, B361987, B406342, B414102, B458076, B468915, B477868, B558538, B562688, B563304, B603078, B654235, B656399, B657059, B666486, B669610,B670343, B677246, B699308, B736874, B743098, B755194, B755196, B755673, B769461, B771317, B775167, B775168; PED3-021-325-12/00847536030602, 00847536035096, 00847536036420/B324297, B407597, B418843, B494876, B561787, B592073, B623579, B629046, B676507, B678997, B716220, B741395, B751748, B787015, B787017; PED3-021-325-14/00847536030619, 00847536035102, 00847536036437/B324809, B347993, B362685, B459598, B558177, B563209, B575964, B580901, B628946, B675724, B678948, B681904, B714342, B714344, B733352, B733353, B769593, B790478; PED3-021-325-16/00847536030626, 00847536035119, 00847536036444/B324954, B339033, B371691, B418451, B460439, B495941, B603378, B604475, B623578, B624422, B660509, B664402, B666503, B667510, B667920, B668523, B680717, B681747, B733912, B739722, B739723, B739724, B746618, B794191, B794192; PED3-021-325-20/00847536030640, 00847536035133, 00847536036468/B325613, B371743, B418905, B421650, B463895, B487451, B495368, B591952, B591955, B644630, B644769, B644809, B657857, B659135, B659138, B659243, B684852, B716733, B716735, B741305, B741316, B741321, B787131; PED3-021-250-12/00847536030404, 00847536034891, 00847536036222/B326236, B335270, B387931, B390988, B394283, B421249, B477292, B497244, B578963, B664457, B666521, B667945, B669726, B696189, B719158, B724199, B724200, B724764, B741102, B746521, B764797, B764798, B787764, B787765, B787766, B793905; PED3-021-300-14/00847536030534, 00847536035027, 00847536036352/B327828, B335957, B346789, B393640, B399293, B403772, B456576, B547554, B568649, B575299, B631125, B631126, B649258, B650123, B652886, B657773, B683152, B711496, B712157, B712158, B732804, B732807, B754755, B754756, B754757, B760951, B760953, B768142, B769947, B789256; PED3-021-250-16/00847536030428, 00847536034914, 00847536036246/B328095, B416673, B452284, B478579, B494048, B565441, B573470, B575302, B628868, B628869, B630965, B656206, B666535, B667323, B668743, B739139, B739405, B739408, B754155, B754886, B754888, B755651, B794406, B794408, B794426; PED3-021-300-12/00847536030527, 00847536035010, 00847536036345/B338969, B345449, B346121, B361988, B405158, B418358, B454923, B551608, B561782, B562698, B563303, B575894, B677108, B677319, B699289, B703245, B718483, B718491, B739905, B744728, B760716, B766588; PED3-021-275-14/00847536030473, 00847536034969, 00847536036291/B343142, B361366, B379727, B429213, B448933, B452378, B461124, B601345, B607550, B624425, B626121, B678023, B680119, B711192, B713287, B724745, B744169, B744175, B744176, B770971, B770988; PED3-021-350-14/00847536030695, 00847536035188, 00847536036512/B349031, B372357, B388981, B391924, B399969, B421758, B445867, B491603, B492310, B497318, B596593, B629649, B630549, B631109, B670495, B678371, B679957, B713023, B713796, B740630, B740631, B742526, B743935, B753618; PED3-021-350-25/00847536030732, 00847536035225, 00847536036550/B364361, B393438, B403082, B419021, B422233, B430531, B437842, B447224, B488266, B616574, B624414, B629444, B629660, B672265, B672892, B673364, B717131, B717132, B724110, B733587, B733590, B741976, B761681, B768780, B771211, B772034, B773456, B773457; PED3-021-350-12/00847536030688, 00847536035171, 00847536036505/B368529, B406425, B419491, B490325, B496666, B497114, B611908, B615889, B616569, B643508, B643509, B644619, B657109, B658427, B659171, B690687, B715610, B715612, B716621, B742378, B746455, B787264; PED3-021-250-10/00847536030398, 00847536036215, 00847536034884/B387475, B390553, B427939, B490270, B575123, B578342, B628755, B654233, B666402, B670376, B693553, B707945, B711113, B740699, B764024, B764025, B766116, B769419, B797931, B801945; PED3-021-275-20/00847536030503, 00847536034990, 00847536036321/B391792, B398628, B418296, B452832, B461730, B497830, B548241, B567772, B655966, B656945, B666505, B671729, B677211, B679094, B680758, B693403, B699179, B711385, B725638, B725639, B736012, B736612, B737225, B751998, B756411, B797986; PED3-021-350-18/00847536036536/B798906; PED3-021-325-18/00847536036451/B800239; PED3-021-275-18/00847536036314/B805162; PED3-021-250-18/00847536036253/B805488; PED3-021-300-18/00847536036376/B807417

Distribution

Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Republic Of Moldova, Montenegro, Morocco, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam

Quantity

7820