FDA Enforcement Class II Terminated

SOMATOM Edge Plus, Model Number 10267000 Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

Recall: Z-1246-2019 · Reported May 8, 2019

Enforcement

Recall Number
Z-1246-2019
Event ID
82227
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 8, 2019
Initiation Date
February 14, 2019
Classification Date
May 2, 2019
Termination Date
May 1, 2020
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

SOMATOM Edge Plus, Model Number 10267000 Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

Reason

Potential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray tube.

Code Info

Serial numbers of all affected models: 122022 100143 100107 83834 83746 83811 66649 122020 83801 122008 122029 83767 83852 100092 83798 100091 100112 100129 100136 83837 122023 83871 100117 100097 100099 122009 83763 83819 83752 83845 83846 83749 83847 100146 83786 83842 83775 100090 83796 83768 83859 100087 83826 83754 83790 122011 83824 100147 83810 83758 100140 100085 83808

Distribution

US Nationwide Distribution

Quantity

53 total units distributed in U.S.