FDA Enforcement Class II Ongoing

Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4. 25GA Laser Probe Illuminated Curved Cat No. VS0125.25B 5. 25GA Laser Probe MaxReach Cat No. VS0140.25 6. 25GA Laser Probe Straight Cat No. VS0110.25

Recall: Z-1245-2026 · Reported February 11, 2026

Enforcement

Recall Number
Z-1245-2026
Event ID
98255
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Vortex Surgical Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 11, 2026
Initiation Date
December 16, 2025
Classification Date
February 4, 2026
Address
4 Research Park Dr Ste 124, N/A, Saint Charles, MO, 63304-5639, United States

Description

Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4. 25GA Laser Probe Illuminated Curved Cat No. VS0125.25B 5. 25GA Laser Probe MaxReach Cat No. VS0140.25 6. 25GA Laser Probe Straight Cat No. VS0110.25

Reason

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Code Info

1. 23GA Laser Probe Curved Pouch UDI 810123480036 Box UDI 810123480180 Lot 2411027 2. 25GA Laser Probe Curved Pouch UDI 810123480043 Box UDI 810123480197 Lot 2411026 3. 25GA Laser Probe Flex-Tip Pouch UDI 810123480111 Box UDI 810123480265 Lot 2411024 2411031 4. 25GA Laser Probe Illuminated Curved Pouch UDI 810123480098 Box UDI 810123480241 Lot 2411025 5. 25GA Laser Probe MaxReach Pouch UDI 810123480135 Box UDI 810123480289 Lot 2410029 2411023 6. 25GA Laser Probe Straight Pouch UDI 810123480012 Box UDI 810123480166 Lot 2411030

Distribution

Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.

Quantity

14,789 (8651 US; 6138 OUS)