FDA Enforcement Class II Ongoing

Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD

Recall: Z-1244-2026 · Reported February 11, 2026

Enforcement

Recall Number
Z-1244-2026
Event ID
98255
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Vortex Surgical Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 11, 2026
Initiation Date
December 16, 2025
Classification Date
February 4, 2026
Address
4 Research Park Dr Ste 124, N/A, Saint Charles, MO, 63304-5639, United States

Description

Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD

Reason

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Code Info

1. Flat Vitrectomy Lens Pouch UDI 810123482290 Box UDI 810123482306 Lots 2410036 2411020 2. Volk Magnifying Lens Pouch UDI 810123482825 Box UDI 810123482832 Lot 2411019

Distribution

Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.

Quantity

14,789 (8651 US; 6138 OUS)