CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 30mm Material Number: 66-230 - Product Usage: intended for anterior cervical fixation (C2-C7) for the following conditions: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudathrosis, and failed previous fusion).
Enforcement
- Recall Number
- Z-1244-2021
- Event ID
- 87251
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- March 17, 2021
- Initiation Date
- January 15, 2021
- Classification Date
- March 11, 2021
- Address
- 375 River Park Cir, Marquette, MI, 49855-1781, United States
Description
CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 30mm Material Number: 66-230 - Product Usage: intended for anterior cervical fixation (C2-C7) for the following conditions: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudathrosis, and failed previous fusion).
Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
Batch Number: 330972 330973 338829 344097 344098 350514 350515 352677 358455 361801 361802 364748 367803 369335 370391 370392 377187 379016 379539 379540 379541
US Nationwide distribution.